Follow us

Timko Dubielzig, Sebastian Halama, Henning Hahn, Giorgio Zarantonello, Malte Niemann, Amado Bautista-Salvador, Christian Ospelkaus

Abstract

We describe the design, commissioning and operation of an ultra-low-vibration closed-cycle cryogenic ion trap apparatus. One hundred lines for low-frequency signals and eight microwave / radio frequency coaxial feed lines offer the possibility of implementing a small-scale ion-trap quantum processor or simulator. With all supply cables attached, more than 1.3 W of cooling power at 5 K is still available for absorbing energy from electrical pulses introduced to control ions. The trap itself is isolated from vibrations induced by the cold head using a helium exchange gas interface. The performance of the vibration isolation system has been characterized using a Michelson interferometer, finding residual vibration amplitudes on the order of 10 nm rms. Trapping of 9Be+ ions has been demonstrated using a combination of laser ablation and photoionization.

Read paper in full

Share article

  • Twitter
  • LinkedIn
English